The compact GL100 temperature humidity datalogger sensor designed for industries requiring climate monitoring and control such as greenhouse, hydroponics, agriculture and botanical rooms gives managers real time environment and climate monitoring. Fine arts museums and galleries are also served for controlled environment settings where the GL100 excels! Remote real-time temperature humidity data monitoring with the GL100 Wi-fi model.

Item
Description
GS-TH
Relative Temperature and Humidity Sensor
 
Measurement Data
Temperature, Humidity, Dew Point Temperature (calculated value)
Accumulated Temperature ºCh
Accumulated Temperature only recorded
Measurement Range
±2G (20m/s2, ±5 (50m/s2, ±10 (100m/s2
(Gravitational Acceleration 1G=approx. 9.807m/s
 
Measuring Method
C-MOS sensor
Range of Temperature Measurement
-20ºC to 85ºC
Temperature Measurement Degree of Accuracy
-20 ≤TS < 0 ±0.8(ºC)
0 ≤TS < 60 ±0.5(ºC)
60 < TS ≤85 ±0.8ºC
Range of Humidity Measurement
0.0 – 100.0% RH
Humidity Measurement Accuracy
At 25ºC
±1.0% RH (0-10% RH)
±8% RH (10-20% RH)
±5% RH (20-80% RH)
±8% RH (80-90% RH)
±10% RH (90-100% RH)

Response Time (Temperature)
Temperature, Humidity: approx. 100 seconds to reach 63% at 25ºC with airflow speed of 1m/s(1)
Processor Functionality
Accumulated Temperature: 0 to 999,999kº ch(2)
Memory Capacity (Memory Mode)
16 | 32 | 64 | 128k sample
Pre-trigger
(Memory Mode)
OFF/10 | 25 | 50 | 100% of selected memory
Sampling Interval
0.5 | 1 | 2 | 5 | 10 | 20 | 30 sec
1 | 2 | 5 | 10 | 20 | 30 | 60 min
Alarm
OFF/level
Temperature Unit of Denomination
ºC/ºF (selectable)
Waterproofing
Waterproof protection filter at the connection point.
 
Operating Environment
Within range of Temp/Humidity measurement.(3)
External Dimensions
15mm x 45mm x 10.2mm (excludes protrusion)
Weight
Approx. 14g (0.49 oz)

(1) Varies by operating environment.
(2) Reset count at start of recording; exceeding or falling below standard temperature.
(3) GL100 main unit??????.