GL860 SPECIFICATIONS
| Item | Description | |
|---|---|---|
| Number of analog input channels | Maximum 200ch available for 1 terminal(20ch/30ch) or extension unit | |
|
External input/
output (*1) |
Input |
Trigger or Sampling (1 channel), Logic/Pulse (4 channels)
Voltage range :0 – +24V(common ground) Input signal: No-voltage contact (a-contact, b-contact, NO, NC) Open collector, Voltage input |
| Output |
Alarm (4 channels)
Output signal: Open collector (pull-up to 5V by 10kΩ resistor) <Maximum rating of the output transistor> Voltage: Max. 30V Current: Max. 0.5A Collector dissipation: Max. 0.2W |
|
| Sampling interval | 5ms – 1h(5ms – 50ms is available for voltage and humidity, Number of channels is limited, external※1 | |
| TIME / DIV | 1 sec to 24 hour | |
| Trigger function | Repeat action | Off, On |
| Trigger type |
Start: Data capture starts when a trigger is generated.
Stop: Data capture stops when a trigger is generated. |
|
| Trigger setting |
(Start / Stop)
Off, Level, Alarm, External, Time, Date, Week |
|
| Alarm function | Alarm judgement mode |
Detection method: Level, Edge
Analog channel judgment mode: H (↑), L (↓), Window In, Window Out Logic channel judgment mode: H (↑), L (↓) Pulse channel judgment mode: H (↑), L (↓), Window In, Window Out Calculation channel judgment mode: H (↑), L (↓), Window In, Window Out |
|
Pulse input
function |
Measurement mode
|
Rotation count (RPM) mode
Accumulating count mode Instant count mode |
| Maximum number of pulse inputs |
Maximum input frequency : 50kHz
Maximum number of count : 50kC/sampling (16-bit counter) |
|
| Interface to PC | USB2.0 (Hi-speed), Ethernet (10BASE-T / 100BASE-TX), Wireless LAN(using B-568 option) | |
|
Storage
device |
Internal | Built-in 8GB Flash Memory (*2) |
| External | One SD card slot (Support SDHC memory card, up to 32GB) ( * 2 ) (*3) | |
| Saved contents | Captured data, Setting conditions, Screen copy | |
| Display | 4.3-inch TFT color LCD (WQVGA: 480 × 272 dots) | |
| Operating environment |
0 to 45 ºC, 5 to 85 % RH (non condensed)
(When operating with battery pack 0 to 40 ºC, charging battery 15 to 35 ºC) |
|
| Power source | AC adapter | 100 to 240 V AC, 50/60 Hz (1 pc of adapter is attached as standard accessory) |
| DC power | 8.5 to 24 V DC (DC drive cable (option B-514) is required) | |
| Battery pack | Mountable battery pack (battery pack (option B-573): 7.2V DC, 2875mAh) | |
| Power consumption |
Max. 36 VA
(Rating under maximum power consumption using the AC adapter, with LCD display on, and battery pack being charged.) |
|
| External dimensions (W x D x H) | Approx.188 x 117 x 42 mm (excluding projections) | |
| Weight | Approx. 500 g (excludes AC adapter and battery pack) | |
GL260 ANALOG INPUT SPECIFICATIONS
| Item | Description | |||
|---|---|---|---|---|
| Type of input terminal | M3 screw type terminals (Rectangular flat washer) | |||
| Input method |
Photo MOS relay scanning system
All channels isolated, balanced input |
|||
| Measurement range | Voltage | 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50, 100 V, 1-5 V F.S. | ||
| Thermocouple | K, J, E, T, R, S, B, N, C (W: WRe5-26) | |||
| Humidity | 0 to 100% (voltage 0 to 1 V scaling conversion) – using the humidity sensor (option B-530) | |||
| Filter | Off, 2, 5, 10, 20, 40 (moving average in selected number) | |||
| A/D converter | Sigma-Delta type, 16 bits (effective resolution: 1/40000 of the measuring full range) | |||
|
Maximum
input voltage |
Between
+/– terminals |
20 mV to 1 V range: 60 Vp-p,
2 V to 100 V range: 110 Vp-p |
||
|
Between
input terminal / input terminal |
60 Vp-p | |||
|
Between
input terminal / GND |
60 Vp-p | |||
Software Specifications for PC
| Item | Description |
|---|---|
| Model name | GL28-APS |
| Supported OS | Windows 11 (64bit) / 10 (64bit/32bit) |
| Functions | Control the GL series, Real-time data capture, and Data format conversion |
| Number of groups | Up to 4 groups |
| Number of CHs per 1 group | Up to number of connected module |
| Maximum number of channels | 1000 ch maximum |
| Settings control | AMP settings, capture settings, trigger/alarm settings, report settings, others |
| Capturing data |
Real-time data (CSV, GBD Binary)
Data in Internal memory or SD memory card (CSV, GBD binary) |
| Display | Analog waveforms, logic waveforms, pulse waveforms, digital values |
| Display modes | Y-T View, Digital View, X-Y View between Cursors (only during replay) |
| File conversion | Between cursors, All data |
| Monitor functions | Alarm monitor enables sending of email to the specified address |
| Statistic/History | Displays maximum, minimum and average values during measurement |
| Report function | Enables creation of daily or monthly files |
WIRELESS LAN UNIT Specifications
| Item | Description |
|---|---|
| Model number | B-568 |
| Communication method | Wireless communication (using radio waves in the 2.4GHz band) |
| Installation |
Attach to the wireless LAN connection terminal
* When the wireless unit is inserted, the SD memory card cannot be inserted into the SD CARD slot. |
| Wireless LAN standard | IEEE802.11b/g/n |
| Function |
Control from PC, data transfer to PC, control and data transfer from smartphone/tablet
• For access point: Local control and transfer are possible. • For station: In addition to local, remote control and transfer using G-REMOTE is also possible. Communication range: Approx. 40 m * Communication range depends on the obstacles and the surrounding environmental conditions. WPS: Push button method / PIN method Encryption function: WEP64, WEP128, WPA-PSK/WPA 2-PSK (TKIP/AES) * WPA/WPA2 and TKIP/AES are automatically selected. |
Humidity Sensor B-530 (option) Specifications
| Item | Description | |
|---|---|---|
| Allowable temperature range | -25 to 80℃ | |
| Allowable humidity range | 0 to 100% RH | |
| Relative humidity measurement accuracy | ±3% RH(5 to 98% RH at 25℃) | |
| Method | Capacitance method | |
| Relative humidity measurement accuracy (5 to 98%) | Measurement environment | Measurement accuracy |
| 0 to 10℃ | ±5% RH | |
| 10 to 20℃ | ±4% RH | |
| 20 to 30℃ | ±3% RH | |
| 30 to 40℃ | ±4% RH | |
| 40 to 50℃ | ±5% RH | |
| 50 to 60℃ | ±6% RH | |
| 60 to 70℃ | ±7% RH | |
| 70 to 80℃ | ±8% RH | |
| * Measurement accuracy at 60℃ or more is a reference value. | ||
| Response time | 15 sec. (90% response when membrance filter is installed) | |
| Sensor output | 0 to 1 VDC | |
| External dimensions | Φ14 x 80 mm (excluding cable) | |
| Cable length | 3 m | |
| Sensor power source | DC + 5 to +16 V | |
| Power consumption | approx. 4 mA | |
OVERVIEW | TERMINALS | G-REMOTE | SOFTWARE | SPECIFICATIONS | ACCESSORIES | DOWNLOAD